Pfeiffer Vacuum Gauge Controller

Pfeiffer Vacuum Gauge Controller

Local
github.comScience & Education23Updated 5w ago

About Pfeiffer Vacuum Gauge Controller

An MCP server for reading and logging pressures from Pfeiffer Vacuum TPG gauge controllers and managing ionisation gauges. It checks pressure before allowing a gauge to be switched on.

Key features

  1. 01Read one or all gauge pressures
  2. 02Log pressure series and save CSV
  3. 03Identify connected gauges
  4. 04Change the pressure unit
  5. 05Switch ionisation gauges with pressure checks
  6. 06Read controller errors

Use cases

  1. 01Monitor chamber and foreline pressure
  2. 02Record pump-down or leak-up pressure curves
  3. 03Identify gauges and their power states
  4. 04Switch on an ionisation gauge after pressure verification

MCP support

Exposes pressure-reading, logging, gauge-identification, and control tools to MCP clients. It connects to supported controllers over USB serial, Ethernet, or RS-232.

Configuration

Local server config is the same for every client. Copy it into your client’s MCP config file.

mcp.json · Universal config
{
"mcpServers": {
"k-dense-ai-labmcp-pfeiffer-tpg": {
"command": "uvx",
"args": [
"labmcp-pfeiffer-tpg"
],
"env": {
"LABMCP_ADDRESS": "<LABMCP_ADDRESS>",
"LABMCP_SIMULATE": "<LABMCP_SIMULATE>",
"LABMCP_READ_ONLY": "<LABMCP_READ_ONLY>",
"LABMCP_LIMITS": "<LABMCP_LIMITS>"
}
}
}
}
Run commanduvx labmcp-pfeiffer-tpg
Environment variables to set
LABMCP_ADDRESSInstrument address, e.g. serial:///dev/ttyUSB0 or tcp://192.168.1.50:5025
LABMCP_SIMULATESet to 1 to use the built-in simulator (no hardware)
LABMCP_READ_ONLYSet to 1 to disable all state-changing tools
LABMCP_LIMITSSafety limit overrides, e.g. max_temperature_c=80

Local servers need a runtime on your machine (e.g. Node.js 18+). Config file locations differ by client; see your client’s docs.

Packages
pypilabmcp-pfeiffer-tpg0.1.2

Tools

11

Saved tool list · updated 2w ago

This tool takes no parameters.

Details

From the project README

Pfeiffer Vacuum Gauge Controller — MCP Server

Let an AI agent read and log chamber and foreline pressures, identify the connected gauges, change the pressure unit and switch ionisation gauges on and off (with a pressure check) on Pfeiffer Vacuum TPG total-pressure gauge controllers.

Packagelabmcp-pfeiffer-tpg
InstrumentsTPG 361 SingleGauge, TPG 362 DualGauge, TPG 366 MaxiGauge (ActiveLine gauges); TPG 261 / TPG 262 (compact gauges)
InterfacesTPG 36x: USB type B (virtual COM port), Ethernet (TCP port 8000); TPG 26x: RS-232
ProtocolPfeiffer mnemonics protocol, ACK/NAK + ENQ (TPG 361/362 communication protocol BG 5510 BEN, TPG 366 communication protocol BG 5511 BEN, TPG 262 operating instructions BG 805 196 BE section 5; originals in the Pfeiffer download center)
Status🧪 simulated: tested against a byte-level simulator of the protocol, not yet verified on hardware. Report a hardware test

Try it without hardware

uvx labmcp-pfeiffer-tpg --simulate --check                      # TPG 362: FullRange gauge + Pirani
uvx labmcp-pfeiffer-tpg --simulate --option model=tpg366 --check # 6 channels incl. a cold cathode gauge

The simulated chamber pumps down from 8e-5 mbar towards 2e-7 mbar (10 min time constant) and the foreline sits at about 2.4e-2 mbar.

Connect your controller

  1. Protocol (TPG 36x): the PROTOCOL parameter must be AUTO (factory setting) or MNE (mnemonics), not PV (Pfeiffer Vacuum telegram protocol).
  2. USB (TPG 36x): the USB type B port appears as a virtual COM port. Factory setting 9600 baud, 8N1 (parameter BAUD USB; 19200-115200 also possible).
  3. Ethernet (TPG 36x, TPG 366): set DHCP or a static IP (front panel or Pfeiffer's Ethernet Configuration Tool); the command port is fixed at 8000.
  4. RS-232 (TPG 261/262): 9600 baud 8N1 by default, only TXD/RXD used.
  5. Test the connection:
    uvx labmcp-pfeiffer-tpg --address /dev/ttyUSB0 --check
    uvx labmcp-pfeiffer-tpg --address COM3 --check
    uvx labmcp-pfeiffer-tpg --address "serial://COM3?baudrate=115200" --check
    uvx labmcp-pfeiffer-tpg --address tcp://192.168.1.70:8000 --check
    
    The model is detected with AYT (TPG 36x); a controller that does not know AYT is treated as a TPG 26x. Use --option model=tpg261 (or tpg262, tpg361, tpg362, tpg366) to be explicit.

Add to your MCP client

Claude Code

claude mcp add vacuum -- uvx labmcp-pfeiffer-tpg --address tcp://192.168.1.70:8000

Claude Desktop / Cursor / Windsurf (claude_desktop_config.json, .cursor/mcp.json, …)

{
  "mcpServers": {
    "vacuum": {
      "command": "uvx",
      "args": ["labmcp-pfeiffer-tpg", "--address", "tcp://192.168.1.70:8000"]
    }
  }
}

Add --read-only to allow reading and logging but block unit changes and switching gauges on (switch_gauge_off stays available). For other clients, generate the snippet with uvx labmcp config pfeiffer-tpg --address tcp://192.168.1.70:8000 --client vscode (also cursor, codex, claude-code).

Tools

ToolKindDescription
get_command_log👁 readReturn the most recent raw commands sent to / replies received from the instrument (newest last). Useful for debugging and for recording what was done.
get_connection_info👁 readReport which instrument is connected (identity, address, simulated or real), whether the server is read-only, and the active safety limits. Call this first.
get_errors👁 readRead (and clear) the controller's ERROR word: controller error, no hardware, inadmissible parameter or syntax error. An empty list means no error.
get_gauge_types👁 readList the gauge connected to each channel (TID), what kind it is, and whether it is an ionisation gauge that is currently switched on or off (SEN).
log_pressure_series👁 readLog pressures at a fixed interval (e.g. a pump-down curve, leak-up/rate-of-rise test or bake-out). Returns per-channel statistics in log-space plus a downsampled series; the full series can be written to a new CSV file (an existing file is never overwritten). Bounded by max_log_duration_s and 2 h per call.
read_all_pressures👁 readRead every channel of the controller at once (PRX), with status per channel.
read_pressure👁 readRead one gauge channel: status, pressure in the display unit and in mbar.
reconnect🛑 safetyClose and re-open the connection to the instrument (e.g. after it was power cycled or a cable was re-plugged).
set_gauge_power⚠️ hazardSwitch an ionisation gauge (IKR, PKR, PBR, IMR) on or off (SEN). Switching a hot or cold cathode gauge on at high pressure can burn out the filament or contaminate the gauge, so switching ON is refused unless a valid reading from this gauge or reference_channel is at or below max_switch_on_pressure_mbar. Switching off may trip setpoint relays used as interlocks.
set_unit🎛 controlChange the pressure unit used on the display and interface (UNI). Readings in mbar are always included regardless of this setting.
switch_gauge_off🛑 safetySwitch an ionisation gauge off (e.g. before venting, or if the pressure is rising). Always allowed. Note that setpoint relays assigned to this channel may change state.

get_connection_info, get_command_log and reconnect are built into every LabMCP server.

Safety limits

LimitDefaultMeaning
max_switch_on_pressure_mbar1e-2 mbarAn ionisation gauge may only be switched on when a valid reading (its own, or reference_channel) is at or below this pressure
max_log_duration_s3600 sLongest log_pressure_series an agent may start

Override at launch: --limit max_switch_on_pressure_mbar=1e-3 --limit max_log_duration_s=7200. One log_pressure_series call logs at most 2 h (the tool's timeout); for a longer bake-out, log in consecutive calls to separate files. save_path must end in .csv and must not exist yet (an existing file is never overwritten); it is checked before logging starts.

Example prompts

  • "What's the chamber pressure right now, and is the foreline below 5e-2 mbar?"
  • "Log all gauges every 10 s for 30 minutes and tell me how many decades per minute the chamber is pumping down."
  • "Close the valve, then log channel 1 every 2 s for 5 minutes so I can estimate the leak-up rate."
  • "Which gauges are connected, and which ones are ionisation gauges that are switched off?"
  • "The chamber is below 1e-5 mbar on the FullRange gauge - switch on the cold cathode gauge on channel 3, using channel 1 as the reference."

Notes

  • Switching ionisation gauges on is the hazard. Hot-cathode (Bayard-Alpert) filaments can burn out and cold-cathode (Penning) gauges contaminate when operated at high pressure. set_gauge_power(on=True) requires a reading with status ok at or below max_switch_on_pressure_mbar from the gauge itself or from reference_channel on the same vacuum. An underrange reading from a Pirani is not accepted as proof (what the controller reports on underrange is not specified), so in a Pirani + Penning setup switch the Penning on at the controller, or use a FullRange gauge as the reference. The 1e-2 mbar default is the upper end of a typical cold cathode measuring range; check your gauge's manual and lower it for hot-filament gauges if recommended.
  • Switching a gauge off (switch_gauge_off, or set_gauge_power(on=False)) can change setpoint (switching-function) relays assigned to that channel, which may be wired to valve or pump interlocks.
  • Every pressure comes with the controller's status: ok, underrange, overrange, sensor error, sensor off, no sensor or identification error. pressure is only filled in when the status is ok; raw_value is what the controller sent. Values are converted to mbar (1 Torr = 1.333224 mbar); in V mode no conversion is made.
  • Every command waits for ACK/NAK; after a NAK the ERROR word (controller error, no hardware, inadmissible parameter, syntax error) is read and reported. get_errors reads and clears it.
  • Degas, filter, calibration factors, switching-function setpoints, gauge control (hot start/self control), the data logger and the Pfeiffer Vacuum (PV) telegram protocol are not exposed. RS-485 multi-drop operation has not been tested.

Hardware verification

ModelFirmwareInterfaceVerified byDate
none yet: be the first

Comments

Markdown supported

New to MCP?

From what the protocol is to installing your first server, these answers cover most first questions.

An open protocol that defines how AI applications connect to external tools and data in a uniform way. Clients talk to the model; MCP servers provide tools, resources and prompt templates. They exchange standard messages, so one server can be reused by many clients.