SRS Residual Gas Analyzer

SRS Residual Gas Analyzer

Local
github.comScience & Education23Updated 5w ago

About SRS Residual Gas Analyzer

SRS Residual Gas Analyzer is an MCP server for SRS RGA100, RGA200, and RGA300 quadrupole mass spectrometers. It measures gas pressures and spectra, supports helium leak checks, and provides first-pass gas identification, with pressure interlocks for the filament, CDEM, and degas operations.

Key features

  1. 01Measure partial and total pressure
  2. 02Record analog and histogram spectra
  3. 03Identify likely gases from major peaks
  4. 04Run bounded helium leak checks
  5. 05Enforce pressure limits for hazardous operations

Use cases

  1. 01Survey residual gases in a vacuum system
  2. 02Monitor selected masses over time
  3. 03Check suspected joints for helium leaks
  4. 04Screen spectra for common gases or contamination

MCP support

Exposes analyzer status, pressure measurements, spectra, gas identification, leak checks, settings, and safety controls as tools for MCP clients.

Configuration

Local server config is the same for every client. Copy it into your client’s MCP config file.

mcp.json · Universal config
{
"mcpServers": {
"k-dense-ai-labmcp-srs-rga": {
"command": "uvx",
"args": [
"labmcp-srs-rga"
],
"env": {
"LABMCP_ADDRESS": "<LABMCP_ADDRESS>",
"LABMCP_SIMULATE": "<LABMCP_SIMULATE>",
"LABMCP_READ_ONLY": "<LABMCP_READ_ONLY>",
"LABMCP_LIMITS": "<LABMCP_LIMITS>"
}
}
}
}
Run commanduvx labmcp-srs-rga
Environment variables to set
LABMCP_ADDRESSInstrument address, e.g. serial:///dev/ttyUSB0 or tcp://192.168.1.50:5025
LABMCP_SIMULATESet to 1 to use the built-in simulator (no hardware)
LABMCP_READ_ONLYSet to 1 to disable all state-changing tools
LABMCP_LIMITSSafety limit overrides, e.g. max_temperature_c=80

Local servers need a runtime on your machine (e.g. Node.js 18+). Config file locations differ by client; see your client’s docs.

Packages
pypilabmcp-srs-rga0.1.1

Tools

18

Saved tool list · updated 2w ago

This tool takes no parameters.

Details

From the project README

SRS Residual Gas Analyzer: MCP Server

Let an AI agent read the residual gas composition of a vacuum system with a Stanford Research Systems RGA100, RGA200 or RGA300 quadrupole mass spectrometer: partial pressures at chosen m/z, analog and histogram spectra, total pressure, helium leak checks and a first-pass identification of the gases (water, hydrogen, nitrogen/CO, oxygen, argon, CO2, hydrocarbons). The filament, the electron multiplier (CDEM) and the ionizer degas are protected by server-side pressure interlocks.

Packagelabmcp-srs-rga
InstrumentsSRS RGA100 (1-100 amu), RGA200 (1-200 amu), RGA300 (1-300 amu), with or without the CDEM option (01)
InterfacesRS-232 (fixed 28,800 baud, 8N1, RTS/CTS), TCP via a serial-to-Ethernet adapter or the SRS RGA Ethernet adapter
ProtocolSRS RGA command set: Operating Manual and Programming Reference, Models RGA100/200/300, Rev. 1.9, chapter 6 "Programming the RGA Head"
Status🧪 simulated: tested against a wire-level simulator, not yet verified on hardware. Report a hardware test

Try it without hardware

uvx labmcp-srs-rga --simulate --check
uvx labmcp-srs-rga --simulate --option scenario=helium_leak    # He pulses every 20 s for leak_check
uvx labmcp-srs-rga --simulate --option scenario=overpressure   # a growing leak trips the filament

The simulator is an RGA200 with a CDEM on an unbaked chamber at ~1e-7 Torr (mostly water, then H2, N2/CO, CO2).

Connect your RGA

  1. Cable: the RGA's electronics box (ECU) has a DB-9 RS-232 port wired as DCE with RTS/CTS hardware handshaking. Use a straight-through cable (or a USB-serial adapter) that carries RTS and CTS. The settings are fixed: 28,800 baud, 8 data bits, no parity, 1 stop bit. The server sets these for you.
  2. Close RGA Windows / RGA Com (only one program can use the port).
  3. Find the port: uvx labmcp ports
  4. Test the connection:
    uvx labmcp-srs-rga --address /dev/ttyUSB0 --check          # Linux
    uvx labmcp-srs-rga --address /dev/tty.usbserial-XXXX --check  # macOS
    uvx labmcp-srs-rga --address COM3 --check                  # Windows
    uvx labmcp-srs-rga --address tcp://192.168.1.50:4001 --check   # raw-TCP serial-to-Ethernet adapter
    uvx labmcp-srs-rga --address tcp://192.168.1.50:818 --option user=admin --option password=admin --check  # SRS RGA Ethernet adapter
    
    On connection the server sends ID? and IN0 (clear the RGA's buffers and run its hardware self-check). It does not switch anything on.

Add to your MCP client

Claude Code

claude mcp add rga -- uvx labmcp-srs-rga --address /dev/ttyUSB0

Claude Desktop / Cursor / Windsurf (claude_desktop_config.json, .cursor/mcp.json, ...)

{
  "mcpServers": {
    "rga": {
      "command": "uvx",
      "args": ["labmcp-srs-rga", "--address", "/dev/ttyUSB0"]
    }
  }
}

Add --read-only to allow measurements but block switching the filament/CDEM, degassing, calibration and settings changes (the off switches stay available). For other clients, generate the snippet with uvx labmcp config srs-rga --address /dev/ttyUSB0 --client vscode (also cursor, codex, claude-code).

Tools

ToolKindDescription
all_off🛑 safetyPut the RGA in a safe state: abort any degas, CDEM off (HV0), filament off (FL0) and quadrupole RF/DC off (MR0). Use when finished, before venting, or if anything looks wrong. It interrupts a running scan or leak check, and every step is attempted even if an earlier one fails.
analog_scan👁 readRecord an analog mass spectrum (SC1): the quadrupole steps through the mass range and the full peak shapes are returned (downsampled) with a peak list, the total pressure measured at the end of the scan, and optionally the full data as CSV. Use it to check peak positions and to survey unknown gases. Duration depends on the noise floor (e.g. 126 ms/amu at NF4). An existing save_path file is never overwritten.
calibrate⚠️ hazardCalibrate the detector: zero (CA) re-zeroes the ion detector at the present noise floor and detector and corrects the RF scan table for temperature drift (seconds); electrometer (CL) recalibrates the electrometer's full I-V response (longer, clears all zero offsets). The quadrupole RF is switched off at the end.
cdem_off🛑 safetySwitch the electron multiplier off (HV0) and return to Faraday-cup detection. Always allowed, and it interrupts a running scan, mass measurement or leak check.
degas⚠️ hazardStart an ionizer degas (DG): 20 mA of 400 eV electrons clean the ion source by electron stimulated desorption. The CDEM is switched off and left off. Degassing shortens filament life; prefer a bakeout. Refused above max_filament_pressure_torr and beyond max_degas_minutes. Returns immediately; the RGA is busy until it finishes (any command would abort it), and filament_off / all_off stop it.
filament_off🛑 safetySwitch the filament off (FL0), stopping a degas first if one is running. Always allowed, and it interrupts a running scan, mass measurement or leak check.
get_command_log👁 readReturn the most recent raw commands sent to / replies received from the instrument (newest last). Useful for debugging and for recording what was done.
get_connection_info👁 readReport which instrument is connected (identity, address, simulated or real), whether the server is read-only, and the active safety limits. Call this first.
get_status👁 readReport the RGA state: filament emission, CDEM on/off and voltage, ionizer settings, scan settings, stored sensitivity factors, and the decoded error bytes (e.g. FL6 when the overpressure protection shut the filament down). Reading the error bytes clears the communication (EC?) and CDEM (EM?) error bytes, as on the instrument.
histogram_scan👁 readRecord a bar-graph spectrum (HS1): one peak-locked value per integer m/z plus the total pressure. Faster than an analog scan and the usual way to follow a residual gas composition.
identify_residual_gases👁 readRun a histogram scan and assign the major peaks to likely gases with a simple lookup of standard 70 eV fragment patterns: H2 (2), He (4), CH4 (16/15), H2O (18/17), N2 or CO (28, split with 14 and 12), O2 (32), Ar (40), CO2 (44), plus diagnoses such as an air leak or hydrocarbon contamination. A screening aid, not a quantitative analysis.
leak_check👁 readHelium leak check: monitor m/z 4 while someone sprays helium on suspect joints, and report the baseline, every response above threshold_factor x baseline (start/end time, peak) and the time series. Keep the first few seconds free of helium so the baseline is clean. The response time of a real leak is a few seconds. Bounded by max_leak_check_duration_s.
measure_masses👁 readMeasure partial pressures at a list of m/z values (peak-locked single-mass measurements, MR). Fast and precise for monitoring known gases; the precision and time per mass are set by the noise floor (set_scan_parameters). Pressures are N2-equivalent.
read_total_pressure👁 readMeasure the total pressure with the RGA acting as an ionisation gauge (TP?, Faraday cup). Needs the filament on and the CDEM off. The value is N2-equivalent and can differ from a Bayard-Alpert gauge by a factor of a few, depending on the gas composition.
reconnect🛑 safetyClose and re-open the connection to the instrument (e.g. after it was power cycled or a cable was re-plugged).
set_cdem⚠️ hazardSwitch the electron multiplier (CDEM) on at a given high voltage for ~100-10000x more signal. High pressure shortens CDEM life and can destroy it, so this is refused unless the pressure is below max_cdem_pressure_torr (RGA total pressure measured now with the Faraday cup, a recent one, or external_pressure_torr), and the voltage is capped by max_cdem_voltage_v. Total-pressure readings are disabled while the CDEM is on.
set_filament⚠️ hazardSwitch the filament on (or change its emission current). The filament burns out or is damaged above 1e-4 Torr, so this is refused unless the pressure is below max_filament_pressure_torr: pass external_pressure_torr from a separate gauge, or, if the filament is already on, the RGA's own total pressure is measured. The RGA's protection will also shut the filament down on overpressure (reported as FL6).
set_scan_parameters🎛 controlChange measurement settings: noise floor (speed vs detection limit), electron energy, ion energy, focus plate voltage and analog-scan resolution. Sensitivity factors stored in the head are only valid at the default ionizer settings (70 eV, 12 eV, 90 V). After changing the noise floor the detector zero is re-adjusted automatically at the start of the next scan.

get_connection_info, get_command_log and reconnect are built into every LabMCP server.

Safety limits

The RGA filament must only run below 1e-4 Torr, and a CDEM should never see voltage above 1e-4 Torr, with 5e-6 Torr or less recommended (manual: FL command warning p. 6-35, "CDEM Handling and Care" in the Maintenance chapter). The RGA only measures pressure while its filament emits, so switching on needs evidence:

  • set_filament and degas: the pressure must be at or below max_filament_pressure_torr. set_cdem: at or below max_cdem_pressure_torr, and the voltage at or below max_cdem_voltage_v (the hardware range of 10-2490 V is enforced as well).
  • Evidence for all three: external_pressure_torr (a reading from another gauge that the user confirmed) and, when the filament is on and the CDEM off, a total-pressure measurement taken right then. If both are available, the higher one is checked, so a mistyped gauge value can't override what the RGA measures. With the CDEM on (the RGA can't measure), a reading from the last max_pressure_reading_age_s is accepted, but only while the filament is still on: once it is off (switched off or tripped) an older reading says nothing about the present pressure.
  • RGA readings used as evidence are scaled to 1 mA emission: the ion current is proportional to the emission current, so a reading at 0.1 mA would otherwise understate the pressure 10x.
  • The instrument's own filament protection still applies: an overpressure shuts the emission and the CDEM down (error FL6), and get_status reports it.
  • all_off, filament_off and cdem_off interrupt a running scan, mass measurement or leak check (which then returns an error and no data) instead of waiting for it to finish. all_off attempts every step even if one fails, and reports what failed.
LimitDefaultMeaning
max_filament_pressure_torr1e-4 TorrHighest pressure at which the filament may be switched on or the ionizer degassed
max_cdem_pressure_torr5e-6 TorrHighest pressure at which the CDEM may be switched on
max_cdem_voltage_v2000 VHighest CDEM high voltage (hardware maximum 2490 V)
max_pressure_reading_age_s60 sOldest RGA total-pressure reading accepted as evidence (only while the filament stays on)
max_degas_minutes3 minLongest ionizer degas (hardware maximum 20 min)
max_scan_duration_s900 sLongest estimated analog/histogram scan
max_leak_check_duration_s1800 sLongest helium leak-check run

Override at launch, e.g. --limit max_cdem_voltage_v=2400 --limit max_filament_pressure_torr=1e-5.

Example prompts

  • "The ion gauge reads 2e-7 Torr. Switch the RGA filament on and tell me what's in the chamber."
  • "Take a 1-100 amu analog scan, save it to ~/data/rga/after_bake.csv, and compare water and hydrogen with this morning."
  • "Is this an air leak? Check 28, 32 and 40 and the ratios."
  • "Start a 3-minute helium leak check on m/z 4; I'll spray the flanges one by one after 20 seconds."
  • "Switch the CDEM on at 1400 V, measure 2, 4, 18, 28, 44 every minute for 10 minutes, then switch everything off."

Notes

  • Wire format. Commands are two letters plus a parameter, ending with CR; ASCII replies end with LF CR. Ion currents from MR, TP? and every scan point are binary 4-byte little-endian two's complement integers in units of 1e-16 A. Hardware commands (IN, FL, HV, EE, IE, VF, CA, CL, DG) answer with the STATUS byte; NF, MI, MF, SA, TP0/1, MR0 answer nothing. The server reads back every silent setting.
  • Rejected commands get no reply. The RGA only flashes its error LED and sets RS232_ERR; the server then reads EC? and reports the reason (bad command, bad parameter, parameter conflict).
  • Partial pressures are N2-equivalent. The server divides the ion current by the partial-pressure sensitivity SP stored in the head (factory value: N2 at default ionizer settings), and by the stored CDEM gain MG when the CDEM is on. Other gases have different sensitivities (H2 ~0.44, He ~0.14, Ar ~1.2 relative to N2); identify_residual_gases corrects with typical values, the other tools do not. The stored factors are only valid at 70 eV / 12 eV / 90 V and 1 mA emission.
  • Total pressure (TP?) is disabled by the RGA while the CDEM is on (it returns 0), and needs the filament on.
  • Degas runs inside the RGA and any command aborts it, so degas returns immediately and the server refuses other commands (except filament_off, cdem_off and all_off, which abort it) until it ends.
  • Noise floor (NF0-NF7) sets the speed and detection limit: from 2000 ms/amu and ~7e-15 A noise (NF0) to 15 ms/amu and ~5e-13 A (NF7). The factory default is NF4 (126 ms/amu).
  • SRS RGA Ethernet adapter login (--option user=... --option password=..., port 818) follows SRS's srsinst.rga driver and has not been tested on hardware. Generic raw-TCP serial servers need no login but must be set to 28,800 baud 8N1 with RTS/CTS.
  • After a scan or measure_masses the server switches the quadrupole RF off (MR0), as the manual recommends.
  • save_path (scans and leak_check) must end in .csv and must not exist yet: existing files are never overwritten. Missing folders are created, ~ is expanded, and the path is checked before the measurement starts.
  • Link recovery. If a scan's binary data doesn't arrive in time (or a safety tool interrupts it), the server discards the partial data, halts the scan with IN0 and waits for its STATUS reply before the next command, so later replies stay in step.

Hardware verification