About risk-management-specialist

An ISO 14971:2019 lifecycle risk management guide for medical-device teams. It organizes planning, analysis, evaluation, control, and post-production monitoring into workflows, with decision frameworks and template references.

Key features

  1. 01Create risk management plans
  2. 02Identify hazards and estimate risks
  3. 03Evaluate risk acceptability
  4. 04Implement and verify risk controls
  5. 05Monitor post-production information
  6. 06Use assessment templates and decision frameworks

Use cases

  1. 01Plan lifecycle risk management for a medical device
  2. 02Analyze hazards and choose an assessment method
  3. 03Document risk controls and verification results
  4. 04Update risk files from complaints and incidents

Install

Copy the command and run it in a terminal.

$npx skills add https://github.com/alirezarezvani/claude-skills --skill risk-management-specialist

After installing, type/risk-management-specialistin Claude Code to confirm it loaded.

Bundled files

References and scripts installed along with SKILL.md.

references/
iso14971-implementation-guide.md
risk-analysis-methods.md
risk-assessment-templates.md
scripts/
fmea_analyzer.py
risk_matrix_calculator.py

Details

From the project README

Risk Management Specialist

ISO 14971:2019 risk management implementation throughout the medical device lifecycle.


Table of Contents


Risk Management Planning Workflow

Establish risk management process per ISO 14971.

Workflow: Create Risk Management Plan
  1. Define scope of risk management activities:
    • Medical device identification
    • Lifecycle stages covered
    • Applicable standards and regulations
  2. Establish risk acceptability criteria:
    • Define probability categories (P1-P5)
    • Define severity categories (S1-S5)
    • Create risk matrix with acceptance thresholds
  3. Assign responsibilities:
    • Risk management lead
    • Subject matter experts
    • Approval authorities
  4. Define verification activities:
    • Methods for control verification
    • Acceptance criteria
  5. Plan production and post-production activities:
    • Information sources
    • Review triggers
    • Update procedures
  6. Obtain plan approval
  7. Establish risk management file
  8. Validation: Plan approved; acceptability criteria defined; responsibilities assigned; file established
Risk Management Plan Content
SectionContentEvidence
ScopeDevice and lifecycle coverageScope statement
CriteriaRisk acceptability matrixRisk matrix document
ResponsibilitiesRoles and authoritiesRACI chart
VerificationMethods and acceptanceVerification plan
Production/Post-ProductionMonitoring activitiesSurveillance plan
Risk Acceptability Matrix (5x5)
Probability \ SeverityNegligibleMinorSeriousCriticalCatastrophic
Frequent (P5)MediumHighHighUnacceptableUnacceptable
Probable (P4)MediumMediumHighHighUnacceptable
Occasional (P3)LowMediumMediumHighHigh
Remote (P2)LowLowMediumMediumHigh
Improbable (P1)LowLowLowMediumMedium
Risk Level Actions
LevelAcceptableAction Required
LowYesDocument and accept; still reduce as far as possible (EU MDR)
MediumAfter reduction AFAPReduce as far as possible; document why further reduction is impossible
HighAfter reduction AFAPReduction required; demonstrate all further options exhausted
UnacceptableNoDesign change mandatory

EU MDR — AFAP, not ALARP: For CE-marked devices, risks must be reduced as far as possible (AFAP) without economic considerations (MDR Annex I, GSPR 1–4; EN ISO 14971:2019/A11:2021 Z-annexes deviation). ALARP ("as low as reasonably practicable"), which permits cost-benefit weighing in acceptability decisions, is not an acceptable criterion under the EU MDR — a notified body will flag it. ISO 14971:2019 itself removed ALARP from the normative text. ALARP may persist in some non-EU jurisdictions (e.g., the UK HSE tradition); if used outside the EU, flag the deviation from EU requirements explicitly.


Risk Analysis Workflow

Identify hazards and estimate risks systematically.

Workflow: Conduct Risk Analysis
  1. Define intended use and reasonably foreseeable misuse:
    • Medical indication
    • Patient population
    • User population
    • Use environment
  2. Select analysis method(s):
    • FMEA for component/function analysis
    • FTA for system-level analysis
    • HAZOP for process deviations
    • Use Error Analysis for user interaction
  3. Identify hazards by category:
    • Energy hazards (electrical, mechanical, thermal)
    • Biological hazards (bioburden, biocompatibility)
    • Chemical hazards (residues, leachables)
    • Operational hazards (software, use errors)
  4. Determine hazardous situations:
    • Sequence of events
    • Foreseeable misuse scenarios
    • Single fault conditions
  5. Estimate probability of harm (P1-P5)
  6. Estimate severity of harm (S1-S5)
  7. Document in hazard analysis worksheet
  8. Validation: All hazard categories addressed; all hazards documented; probability and severity assigned
Hazard Categories Checklist
CategoryExamplesAnalyzed
ElectricalShock, burns, interference☐
MechanicalCrushing, cutting, entrapment☐
ThermalBurns, tissue damage☐
RadiationIonizing, non-ionizing☐
BiologicalInfection, biocompatibility☐
ChemicalToxicity, irritation☐
SoftwareIncorrect output, timing☐
Use ErrorMisuse, perception, cognition☐
EnvironmentEMC, mechanical stress☐
Analysis Method Selection
SituationRecommended Method
Component failuresFMEA
System-level failureFTA
Process deviationsHAZOP
User interactionUse Error Analysis
Software behaviorSoftware FMEA
Early design phasePHA
Probability Criteria
LevelNameDescriptionFrequency
P5FrequentExpected to occur>10⁻³
P4ProbableLikely to occur10⁻³ to 10⁻⁴
P3OccasionalMay occur10⁻⁴ to 10⁻⁵
P2RemoteUnlikely10⁻⁵ to 10⁻⁶
P1ImprobableVery unlikely<10⁻⁶
Severity Criteria
LevelNameDescriptionHarm
S5CatastrophicDeathDeath
S4CriticalPermanent impairmentIrreversible injury
S3SeriousInjury requiring interventionReversible injury
S2MinorTemporary discomfortNo treatment needed
S1NegligibleInconvenienceNo injury

See: references/risk-analysis-methods.md


Risk Evaluation Workflow

Evaluate risks against acceptability criteria.

Workflow: Evaluate Identified Risks
  1. Calculate initial risk level from probability × severity
  2. Compare to risk acceptability criteria
  3. For each risk, determine:
    • Acceptable: Document and accept (EU MDR: still reduce as far as possible)
    • Reduction required (AFAP): Proceed to risk control
    • Unacceptable: Mandatory risk control
  4. Document evaluation rationale
  5. Identify risks requiring benefit-risk analysis
  6. Complete benefit-risk analysis if applicable
  7. Compile risk evaluation summary
  8. Validation: All risks evaluated; acceptability determined; rationale documented
Risk Evaluation Decision Tree
Risk Estimated
      │
      ▼
Apply Acceptability Criteria
      │
      ├── Low Risk ──────────► Accept and document
      │
      ├── Medium Risk ───────► Reduce as far as possible (AFAP)
      │   │                    Document why further reduction impossible
      │   ▼
      │   Further reduction possible?
      │   │
      │   Yes──► Implement control
      │   No───► Document AFAP rationale (no economic considerations)
      │
      ├── High Risk ─────────► Risk reduction required
      │   │                    Must demonstrate reduction AFAP
      │   ▼
      │   Implement control
      │   Verify residual risk
      │
      └── Unacceptable ──────► Design change mandatory
                               Cannot proceed without control
AFAP Demonstration Requirements (EU MDR)
CriterionEvidence Required
All control options consideredAnalysis of every feasible control per the hierarchy (design, protective measures, information)
Further reduction impossibleEvidence each remaining option is technically infeasible or does not further reduce risk
State of the artComparison to similar devices and current standards
Stakeholder inputClinical/user perspectives

Economic considerations (cost of further risk reduction) must not enter the EU acceptability decision (MDR Annex I GSPR 2; EN ISO 14971:2019/A11:2021). Cost may inform business decisions about whether to market the device — never whether a risk is acceptable.

Benefit-Risk Analysis Triggers
SituationBenefit-Risk Required
Residual risk remains highYes
No feasible risk reductionYes
Novel deviceYes
Unacceptable risk with clinical benefitYes
All risks lowNo

Risk Control Workflow

Implement and verify risk control measures.

Workflow: Implement Risk Controls
  1. Identify risk control options:
    • Inherent safety by design (Priority 1)
    • Protective measures in device (Priority 2)
    • Information for safety (Priority 3)
  2. Select optimal control following hierarchy
  3. Analyze control for new hazards introduced
  4. Document control in design requirements
  5. Implement control in design
  6. Develop verification protocol
  7. Execute verification and document results
  8. Evaluate residual risk with control in place
  9. Validation: Control implemented; verification passed; residual risk acceptable; no unaddressed new hazards
Risk Control Hierarchy
PriorityControl TypeExamplesEffectiveness
1Inherent SafetyEliminate hazard, fail-safe designHighest
2Protective MeasuresGuards, alarms, automatic shutdownHigh
3InformationWarnings, training, IFULower
Risk Control Option Analysis Template
RISK CONTROL OPTION ANALYSIS

Hazard ID: H-[XXX]
Hazard: [Description]
Initial Risk: P[X] × S[X] = [Level]

OPTIONS CONSIDERED:
| Option | Control Type | New Hazards | Feasibility | Selected |
|--------|--------------|-------------|-------------|----------|
| 1 | [Type] | [Yes/No] | [H/M/L] | [Yes/No] |
| 2 | [Type] | [Yes/No] | [H/M/L] | [Yes/No] |

SELECTED CONTROL: Option [X]
Rationale: [Justification for selection]

IMPLEMENTATION:
- Requirement: [REQ-XXX]
- Design Document: [Reference]

VERIFICATION:
- Method: [Test/Analysis/Review]
- Protocol: [Reference]
- Acceptance Criteria: [Criteria]
Risk Control Verification Methods
MethodWhen to UseEvidence
TestQuantifiable performanceTest report
InspectionPhysical presenceInspection record
AnalysisDesign calculationAnalysis report
ReviewDocumentation checkReview record
Residual Risk Evaluation
After ControlAction
AcceptableDocument, proceed
Reduced AFAPDocument rationale (no economic considerations), proceed
Still unacceptableAdditional control or design change
New hazard introducedAnalyze and control new hazard

Post-Production Risk Management

Monitor and update risk management throughout product lifecycle.

Workflow: Post-Production Risk Monitoring
  1. Identify information sources:
    • Customer complaints
    • Service reports
    • Vigilance/adverse events
    • Literature monitoring
    • Clinical studies
  2. Establish collection procedures
  3. Define review triggers:
    • New hazard identified
    • Increased frequency of known hazard
    • Serious incident
    • Regulatory feedback
  4. Analyze incoming information for risk relevance
  5. Update risk management file as needed
  6. Communicate significant findings
  7. Conduct periodic risk management review
  8. Validation: Information sources monitored; file current; reviews completed per schedule
Information Sources
SourceInformation TypeReview Frequency
ComplaintsUse issues, failuresContinuous
ServiceField failures, repairsMonthly
VigilanceSerious incidentsImmediate
LiteratureSimilar device issuesQuarterly
RegulatoryAuthority feedbackAs received
ClinicalPMCF dataPer plan
Risk Management File Update Triggers
TriggerResponse TimeAction
Serious incidentImmediateFull risk review
New hazard identified30 daysRisk analysis update
Trend increase60 daysTrend analysis
Design changeBefore implementationImpact assessment
Standards updatePer transition periodGap analysis
Periodic Review Requirements
Review ElementFrequency
Risk management file completenessAnnual
Risk control effectivenessAnnual
Post-market information analysisQuarterly
Risk-benefit conclusionsAnnual or on new data

Risk Assessment Templates

→ See references/risk-assessment-templates.md for details

Decision Frameworks

Risk Control Selection
What is the risk level?
        │
        ├── Unacceptable ──► Can hazard be eliminated?
        │                    │
        │                Yes─┴─No
        │                 │     │
        │                 ▼     ▼
        │            Eliminate  Can protective
        │            hazard     measure reduce?
        │                           │
        │                       Yes─┴─No
        │                        │     │
        │                        ▼     ▼
        │                   Add       Add warning
        │                   protection + training
        │
        └── High/Medium ──► Apply hierarchy
                            starting at Level 1
New Hazard Analysis
QuestionIf YesIf No
Does control introduce new hazard?Analyze new hazardProceed
Is new risk higher than original?Reject control optionAcceptable trade-off
Can new hazard be controlled?Add controlReject control option
Risk Acceptability Decision
ConditionDecision
All risks LowAcceptable
Medium risks reduced AFAPAcceptable
High risks reduced AFAP, documentedAcceptable if benefits outweigh
Any Unacceptable residualNot acceptable - redesign

Tools and References

Scripts
ToolPurposeUsage
risk_matrix_calculator.pyCalculate risk levels and FMEA RPNpython risk_matrix_calculator.py --help

Risk Matrix Calculator Features:

  • ISO 14971 5x5 risk matrix calculation
  • FMEA RPN (Risk Priority Number) calculation
  • Interactive mode for guided assessment
  • Display risk criteria definitions
  • JSON output for integration
References
DocumentContent
iso14971-implementation-guide.mdComplete ISO 14971:2019 implementation with templates
risk-analysis-methods.mdFMEA, FTA, HAZOP, Use Error Analysis methods
Quick Reference: ISO 14971 Process
StageKey ActivitiesOutput
PlanningDefine scope, criteria, responsibilitiesRisk Management Plan
AnalysisIdentify hazards, estimate riskHazard Analysis
EvaluationCompare to criteria, AFAP assessment (EU)Risk Evaluation
ControlImplement hierarchy, verifyRisk Control Records
ResidualOverall assessment, benefit-riskRisk Management Report